Diagnosis and maintenance of common fault phenomena of atomic absorption spectrophotometer

The atomic absorption spectrophotometer is a specialized analytical instrument that utilizes atomic absorption spectroscopy to determine the concentration of specific elements in a sample. The basic principle involves atomizing the sample and measuring the absorption of light at specific wavelengths corresponding to the element being analyzed. This technique is widely used in environmental, pharmaceutical, and industrial laboratories for precise elemental analysis. Common malfunctions of atomic absorption spectrophotometers and their troubleshooting and maintenance methods are outlined below for reference by technical personnel responsible for instrument care. Symptom 1: The display does not light up, and the fan does not start. First, check whether the fuse is loose or has blown. If the fuse is faulty, replace it with a new one of the same rating. Next, inspect the power cord for any damage, open circuits, or poor connections. Ensure that the power outlet is functioning properly. In addition, verify that the internal power supply unit is receiving voltage. It's also important to check the insulation resistance between the power line and the chassis. Normally, this resistance should be greater than 10 MΩ to ensure safety and proper operation. If these steps do not resolve the issue, further inspection of the power board or control system may be necessary. Always follow the manufacturer’s guidelines and safety procedures when performing any maintenance.